01Make the comparison fair
Confirm both instruments' ranges, calibration status, units and concentration basis. Identify whether they sample the same location at the same time. Different sample pressure or a long transport delay can create an apparent disagreement during a changing process.
02Review the sample arrangements
Check materials, connection integrity, flow and stabilization with the relevant suppliers. Trace-oxygen measurements are particularly sensitive to sample contamination. A portable analyser connected temporarily should not automatically override an installed instrument without checking its setup.
03Resolve with a defined method
Use an agreed reference or verification plan and preserve all results, including unfavourable ones. Decide how process-quality release is managed while the issue remains unresolved. After correction, document which cause was established and how the installed readings were confirmed.