01Confirm the measurement basis
Record sensor location, sample pressure, displayed units and instrument status. Check whether a recent configuration change switched between pressure and atmospheric dew point. Compare only readings expressed on the same basis.
02Trace the treatment conditions
Review dryer inlet temperature and flow, regeneration status, drain behaviour and recent maintenance. A moisture spike during a high-temperature peak suggests a different investigation from a persistent abnormal value immediately after sensor work.
03Preserve quality control
Follow the approved response to unsuitable inlet or product moisture. Have qualified personnel correct the identified issue and verify the downstream condition. A cleared alarm alone does not establish that affected equipment or the process gas has returned to the required state.
These public references explain general technology. They do not establish AIDEKE model performance, a supply commitment or a third-party endorsement. The article is an independent editorial explanation; the confirmed technical agreement governs a specific project.